Operator’s Manual
EAR99 technology subject to restrictions on copyrights page.
3
•
Do not use the probe with accessories other than those with which it is shipped.
•
Do not use the probe other than as specified in this manual.
•
Do not use the probe if it is damaged.
Safety Ratings for Accessories
Accessory
Part Number
CAT Rating*
Max. Input Voltage Probe &
Accessory (input to ground)
1x Attenuating Tip
HVFO100-1X-TIP-U
No Rated Measurement Category
+/-1 V (DC + Peak AC)
(5 V maximum non-destruct)
5x Attenuating Tip
HVFO100-5X-TIP-U
No Rated Measurement Category
+/-5 V (DC + Peak AC)
(25 V maximum non-destruct)
10x Attenuating Tip HVFO100-10X-TIP-U
No Rated Measurement Category
+/-10 V (DC + Peak AC)
(50 V maximum non-destruct)
20x Attenuating Tip HVFO100-20X-TIP-U
No Rated Measurement Category
+/-20 V (DC + Peak AC)
(100 V maximum non-destruct)
40x Attenuating Tip HVFO100-40X-TIP-U
No Rated Measurement Category
+/-40 V (DC + Peak AC)
(100 V maximum non-destruct)
* Per IEC/EN 61010-031:2015 definitions, No Rated Measurement Category applies to other circuits that are not directly
connected to the mains supply.
WARNING:
The probe and its accessories are not rated for measurements within
Measurement Categories II, III or IV.