Optical Modulation Analyzer
22
Preparing for Optimal Measurements
The following procedures will prepare the oscilloscope to yield the most accurate measurements. We
highly recommend performing each of them before starting an OMA test session.
Warm Up
Power on all components and allow them to warm for at least 20 min. before starting a test session.
Deskew Channels
The LabMaster oscilloscope has several self-calibration functions to correct for skew between channels.
ChannelSync Calibration should be performed when the LabMaster is set up for the first time and
thereafter whenever a new acquisition module is installed.
NOTE: You will need a high-bandwidth RF cable to connect from the Fast Edge output of the MCM-Zi-A to
each of the input channels on the acquisition module.
1.
From the menu bar, choose Vertical > Channel 1 Setup and click the ChannelSync Cal button.
2.
On the ChannelSync Calibration Wizard, choose the input configuration in use: Input A (four
standard inputs) or Input B (two DBI inputs).
3.
Set V/Div to 50 mV and touch Next.
4.
Follow the prompts to connect the cable from the fast edge output to each channel in turn.
5.
When the calibration is finished, select and apply all the calibration values on the ChannelSync Cal
dialog.