Optical Modulation Analyzer
26
Optical Inputs
Touch the Optical Inputs workflow button to access the OMA’s optical configuration dialogs.
Optical Inputs
This subdialog is used to select the optical level parameters of the signal under test.
NOTE: The OMA software will perform a calibration every time it detects changes in its settings. However,
if there are changes on the transmitter side (such as output power change), then a manual calibration
(Force Calibration) may be required for better signal characterization.
Control
Action
2-Pol
Specifies signal is polarization multiplexed.
1-Pol
Specifies signal is not polarization multiplexed.
Force Calibration
Repeats calibration.
Modulation Type
Sets signal’s digital modulation scheme. Select from a wide variety of
complex modulation formats, or use a Custom format.
Symbol Rate
Sets signal’s symbol rate in bauds.
Laser Source
Specifies the Local Oscillator source: coherent receiver Internal LO output
or External source.