4.3.6 Version Screen
Move the ◄►arrow key to More and press
Enter
. With
Version
blinking, press
Enter
. The screen displays the manufacturer, model, and
software version information.
4.3.7 Showing Negative Oxygen Readings
For software version 1.4.4 or later, the instrument only displays
oxygen readings that are positive or zero. The instrument can be
reconfigured to show negative readings if sensor output drifts below
zero. This situation may arise after the instrument has been zeroed, as
time progresses the sensor may drift below the zero calibration setpoint.
To show negative oxygen readings on the display:
Press the
System
key.
TRAK/HLD Auto-Cal
PSWD Logout More
Use the Right or Left arrow keys and select More. Press
Enter
.
Version Self-Test
Show_Negative=NO
Use the ◄► arrow keys and select “
Show_Negative=NO
”.
Use the ▲/▼ keys to toggle from
NO
to
YES
.
Press the
Escape
key twice to return to the analyze mode.
This preference is stored in non-volatile memory, so this
configuration is remembered after a power shutdown. If the instrument
is cold started, it will go back to default (not showing negative oxygen
readings).