Programming Manual
–
T3AWG3K-C Series Arbitrary Waveform Generator
True-Arb Operating Mode
142
This example generates a sequence of waveforms in Advanced Mode.
In this mode the execution of the sequence can be changed by using conditional and
unconditional jumps (JUMPTO and GOTO commands) and dynamic jumps (PATTERN JUMP
commands).
The sequencer by default is loaded as follows:
Entry 1
–
Length 16384 -
Wait Event Button | Rep. 100
|Infinite Repetitions Off |
Jump Event None | Jump To
Mode Next|Jump to Entry
Item 1|
Go To Mode Next
|
Go To Entry Item 1 | Pattern
Mode Next | Pattern Code
1|Pattern Jump To Entry Item
1
Entry 2
–
Length 16384 -
Wait Event None | Rep. 1
|
Infinite Repetitions On
|
Jump Event Timer
|
Jump To
Mode Next|Jump to Entry
Item 3|
Go To Mode Next |
Go To Entry Item 1 | Pattern
Mode Next | Pattern Code
1|Pattern Jump To Entry
Item 1
Entry 3
–
Length 16384
–
Wait Event None | Rep.
1 |
Infinite Repetitions
On |
Jump Event Button |
Jump
To
Mode
Previous
|Jump to Entry
Item 1|Go To Mode
Next | Go To Entry Item
1 |
Pattern Mode Next |
Pattern
Code
123|Pattern Jump To
Entry Item 1
CH1
Sine, Amp. 2V, Offset 0V
Ramp, Amp. 2V, Offset 0V
Sinc, Amp. 2V, Offset
0V
CH2
Sine, Amp. 2V, Offset 0V
Ramp, Amp. 2V, Offset 0V
Sinc, Amp. 2V, Offset
0V
Before running the VI you can configure in the “Digital Settings” the number of digital
channels, the digital pod voltage level and the digital skew. You can also add a Digital
Waveform to the sequencer.
4.
Run the VI. By default the instrument will receive the trigger from the front panel
button.
5.
Press
the “START AWG” button to run the AWG
.
6.
You can p
ress the “SEND TRIGGER” button to send a software trigger
, you can press
the “SEND PATTERN” button to send the pattern code value.
7.
By default the execution of the sequencer is the following one: press the SEND
TRIGGER button to start the waveform repetitions of the first entry.