49391
Revision 3
54
Gasurveyor 700
USER MANUAL
7. If the displayed reading does not match the concentration of the calibration gas, correct by
pressing the appropriate soft-key.
Press
+
button to increase the reading
Press
-
button to decrease reading
Figure 88: Field Calibration - Adjustment
Figure 89: Field Calibration - Completed Adjustment
8. When the gas reading corresponds to the concentration of calibration gas, disconnect the
tubing from the monitor.
9. Press the
Range
button to change range and calibrate another gas. Repeat steps 4 to 9 for
each range.
10. Press the
End
button after all required ranges have been calibrated.
11. Save screen will appear (as shown in
Figure 90: Field Calibration Save
).
Figure 90: Field Calibration Save
Press
Yes
button to save the calibration.
Press
No
button to exit the Field Calibration mode without saving the calibration.
Press
Back
button to return to the Field Calibration mode.