49391
Revision 3
57
Gasurveyor 700
USER MANUAL
10. Monitor Specifications
Gas
Range
Resolution
Sensor Type
Flammable
0-10,000ppm
0-10,000ppm
1ppm
25ppm
Semiconductor
Infrared
Flammable
0 to 9.9% LEL
10 to 100% LEL
0.1%
1%
Infrared
Flammable
0 to 5% Volume
5 to 100% Volume
0.1%
1%
Infrared
Oxygen (O
2
)
0 to 20.9%
21 to 25%
0.1%
1%
Electrochemical
Carbon Monoxide (CO)
0 to 1000ppm
1ppm
Electrochemical
Hydrogen Sulfide (H
2
S)
0 to 100ppm
1ppm
Electrochemical
Other Specifications
Dimensions (excluding probe): 190mm x 98mm x 107mm (7.4” x 3.8” x 4.2”)
Weight:
Alkaline - 1.4 kg (3.1lbs.) / Rechargeable -1.3 kg (2.7 lbs.)
Temperature Limits:
-20°C to 50°C (-4°F to 122°F)
Humidity Limits:
0 – 90% R.H. non-condensing
Protection Rating:
Polycarbonate / ABS case with TPE over-moulding
protected to IP55
Sampling System:
Integral pump with flow fail sensor. Sample path is protected
by a hydrophobic and cotton filter.
Flow rate: ~ 0.5 l/min with no restriction.
Response times increase ~2 second per metre of tubing used.
Maximum tubing length 30 metres.
Power Source:
Alkaline Batteries: 3 x size ‘D’ (LR20), or Rechargeable
Battery Pack.
See
for list of approved alkaline
batteries.
Battery Life:
Alkaline: 15 hours
Rechargeable: 20 hours
Alarms:
360° highly visible flashing LED
Top panel highly visible flashing LED
Sounder ~ 85 db at 0.3 m
Display:
Monochrome LCD (240 x 160 pixels) with automatic bi-colour
backlight