DPN 402197 Issue 7.1
© Teledyne Limited
D- 2
This information can be recorded on the Configuration Log Form in
.
NOTE
To access the 350 System Configuration dialog press [SHIFT+F2]
TSS supply the 350 system with port and starboard coil triads assembled and the coil calibration
constants already configured in the topside computer, as shown in
the values displayed in 350 System Parameters Configuration screen correspond to the values
stamped on the brass connector flanges of the search coils.
Figure D-1: 350 System Parameters Configuration screen
If a search coil is replaced, the new 5-digit value for the calibration constant must be entered for the
relevant search coil. This will not affect the operation of any of the other remaining search coils.
Each of the six coil calibration constants will be different and ensure they are entered correctly. The
numbers include an error-checking element helping to ensure valid data entry.
When the calibration constants have been correctly entered into DeepView, click OK to accept the
configuration. If the parameters have changed, the new values will be downloaded to the SEP.
D.2 Operation
The Coil Tester is supplied with default settings of 25Hz. To change the frequency specified, see
.