D-3
© Teledyne Limited
DPN 402197 Issue 7.1
1. Power on the 350 system and ensure the DeepView is running to confirm the test results.
Ensure the coil calibration constants have been entered correctly and the 350 system is setup
correctly (see
2. Select the Terminal Window into view and set focus to this window. Press [CTRL+U]. This will
display a warning message stating that continuing will stop communication with the SEP. Click
‘YES’ and the 350 Terminal Menu will be displayed.
3. Press [2] to access the Pre-amp Gain Menu. There are two fields on this screen outlined in
below.
NOTE
To cycle through the available options press [SPACE] bar. Press [ENTER] to
accept settings.
4. As outlined above, set the Pre-amp Gain to 4, Pre-amp Autogain to OFF.
5. When the pre-amp gain and autogain have been set, a chevron will be displayed on screen.
6. Press [4] to enter the Data Window/FFT Size Menu. Set the tone frequency to 25Hz and press
[ENTER] to accept the selection and return to the Terminal Menu.
7. Press [9] to exit the Terminal menu.
8. To re-establish communication with the 350 system, press the Enable/disable polling on the
Terminal Window toolbar. Scrolling data will be displayed on screen to identify the system is
operating correctly.
9. Place the Coil Tester over the end of the search coil so the end of the coil is seated fully into
the circular recess of the tester.
10. Note which coil is being tested. This will be defined on the attached connector cable, as out-
lined in
Table D-1: 350 System Subsea Parameters
Parameter
Options
Required Setting
Pre-amp Gain
1,2,3 and 4
4
Pre-amp Autogain
ON or OFF
OFF
Table D-2: 350 System Connector Cable Identification
Connector ID
Description
SV
Starboard Vertical
SL
Starboard Lateral
PV
Port Vertical
PL
Port Lateral