*To test the port or starboard fore/aft coils, the forward search must be used
11. On the Coil Tester press and hold the circular power switch. The battery condition indicator
LED should first show orange as the unit performs a self test and then, approximately one sec-
ond later, shows green. If the LED goes red or doesn’t illuminate at all, this indicates that the
battery power is low. If necessary, follow the instructions in
to replace the bat-
tery.
12. While continuing to press the power switch, check DeepView and confirm that the channel
being tested shows a signal strength of 1.0 to 1.5e6. The digital display at the bottom left hand
corner of the Run Window should show a value between 1e6 and 1.5e6.
13. Open the Spectrum Analyser to examine the raw signals received by the six search coil chan-
nels. The horizontal axis of the display represents the frequency. The default view is 25Hz. It is
recommended that the frequency scale is raised to 50Hz to view the results. The logarithmic
vertical axis shows the absolute magnitude of signals at each frequency within the displayed
band. This screen will confirm that the tested channel is receiving the tone at a frequency of
25Hz. After the results have been confirmed, toggle through the frequency range to return it to
the default scale of 25Hz.
14. If the Run Window does not show the expected values, follow the guidelines included in
15. Repeat steps 4 to 12 for the remaining channels of the 350 system.
16. After completing the coil test procedure for all six channels, restore the 350 system to the orig-
inal tone frequency for the survey. It is important to reset the pre-amp gain and autogain to
default settings. Repeat steps 2 and 3 using the values outlined below.
17. Press [9] to exit the Terminal Menu.
18. To re-establish communication with the 350 system, press the Enable/disable polling button on
the Terminal Window toolbar. Scrolling data will be displayed on screen to identify the system
is operating correctly.
D.2.1 Frequency Selection
Tests can be run at different frequencies ranging from 25Hz to 21Hz. To do this, remove the Coil Tester
end cap and adjust the Frequency Selection Switch.
When supplied by Teledyne TSS, the Coil Tester will be set to its default position five and 25Hz. It is
strongly recommended that a default test is carried out using this frequency setting.
SF
Starboard Fore/Aft
PF
Port Fore/Aft
Table D-3: 350 System Operating Parameters
Parameter
Required Setting
Pre-amp Gain
4
Pre-amp Autogain
ON
Table D-2: 350 System Connector Cable Identification
Connector ID
Description