Operation
LXT-220 DO
Teledyne Analytical Instruments
34
Press both (CALIBRATE)
◄
/
►
keys again to enter a new
operational password. 1 to 8 digits may he entered. Press either of the
menu keys to accept the new password when done.
Note: A new password will not be accepted unless the protection
level has already been set to a value greater than 0.
3.3.1.3 C
HANGING
C
ONFIGURATION
P
ASSWORD
To change the configuration password, use the MENU SELECT
▲
/
▼
keys to navigate to the PASSWORD menu and select it by
pressing both CALIBRATE
◄
/
►
keys simultaneously. Use the MENU
SELECT
▲
/
▼
keys to bring the cursor to the Configure password level
screen.
To enter a new configure password, navigate to the old password
(or no PW) using the (CALIBRATE)
◄
/
►
keys and select by pressing
both (CALIBRATE)
◄
/
►
keys simultaneously. This brings up the new
configure password screen.
Press both (CALIBRATE)
◄
/
►
keys again to enter a new
configure password. 1 to 8 digits may he entered. Press either of the
menu keys to accept the new password when done.
Note: A new configure password will not be accepted unless the
protection level has already been set to 1.
These two displays work the same as the corresponding operational
displays, except that the protection level can only be 0 or 1.
Config password level:
1
no PW
New opr password