6. Does VueLink2.0 include Alarm Mapping features?
Alarm mapping software features are scheduled to be completed in a future update
7. Can MUVE C360 be Decontaminated after a mission?
The Muve C360 Sensor is IP43 rated. Surfaces can be wiped down with a mild detergent. Do Not Submerge. DJI
recommends Dahlgren Decon (First Line Technologies) for the drone
8. Is the MUVE C360 Intrinsically Safe?
No. Drone platforms are not intrinsically safe. However, testing by Southern Manatee County showed that
drones can fly safely in explosive atmospheres.
9. Is the MUVE C360 Calibration Station compatible with other multi-gas detectors?
The Cal Station is proprietary to the C360, but it utilizes standard industry Cal gasses and fittings
10. What if the MUVE C360 doesn’t pass Calibration?
1. Verify that the Calibration gas displayed on the bottle, matches those listed in the C360 User’s Guide
a. If using a different gas concentration than specified, adjust Calibration Station settings to match
what you are using. Settings are accessed through the MENU button.
2. Verify calibration gasses are not expired
3. Check the pressure on the regulator to ensure the canister is not empty.
4. Verify hose connections are in the proper port as described in the quick start guide.
5. Inspect tubing at hose connections for blockages. The interior of the tubing contains a Teflon liner that
can delaminate with repeated use, blocking gas flow.
*Note: If you observe this issue, please contact our Service Team (below) to obtain replacement
tubing. This known issue is permanently resolved in new C360 shipments.
6. Ensure black Calibration Station O-Ring is lubricated and free of cuts or defects. Refer to User Guide for
details. Lubricating oil and replacement O-rings are provided in the maintenance kit.
11. What is the best method for demonstrating the MUVE C360?
Acetone vapor works extremely well with the PID for static demos and in-flight demos. Demonstration video is in
production, showing live operation and gas detection.