OLCT 60
FIXED POINTGAS MONITOR
USER MANUAL
24
NPO60GB
Revision F.2
5.2.1
Frequency of replacement
The sensor block needs to be replaced every time when zeroing, performing gas calibration or
preventive maintenance are no longer possible.
5.2.2
Exchanging of the sensor
Step
Action
1.
Prepare the following elements:
New sensor block
4 mm Allen wrench
Calibration set (gas cylinder, calibration cup, etc.).
2.
Inhibit the alarms on the controller.
3.
Turn the OLCT 60 off.
4.
Loosen the locking screw in the sensor head and rotate the sensor head 30 degrees
counterclockwise.
5.
Unplug the connector and remove the defective sensor head.
6.
Replace the worn out detector head with an identical new one.
7.
Reassemble in reverse order and tighten the locking screw.
8.
Power the OLCT 60.
9.
Install the OLCT 60 as explained in detail in the Initialization of the sensor block
paragraph on page Erreur ! Signet non défini..
5.3
Initialization of the sensor block
5.3.1
Selection of the initializing menu (
Init
)
Step
Action
Illustration
1a.
After the startup phase, the screen will show the gas
measurement (it may be wrong at this point). Place the magnet
over
for 3 seconds.
1b.
Until the
icon is displayed...
...present the magnet 3 consecutive times on the
in the 3
seconds.
1c.
The calibration menu (CAL) is displayed.
1d.
Place the magnet over
.
Summary of Contents for OLCT 60
Page 1: ...NPO60GB Revision F 2 OLCT 60 FIXED POINTGAS MONITOR USER MANUAL ...
Page 44: ...OLCT 60 FIXED POINTGAS MONITOR USER MANUAL 38 NPO60GB Revision F 2 ...
Page 50: ...OLCT 60 FIXED POINTGAS MONITOR USER MANUAL 44 NPO60GB Revision F 2 ...
Page 52: ...OLCT 60 FIXED POINTGAS MONITOR USER MANUAL 46 NPO60GB Revision F 2 ...
Page 66: ...OLCT 60 FIXED POINTGAS MONITOR USER MANUAL 60 NPO60GB Revision F 2 ...
Page 67: ...OLCT 60 FIXED POINTGAS MONITOR USER MANUAL NPO60GB Revision F 2 61 ...