Page 21 of 42
After
SETUP/CAL
is selected and a channel number has been chosen, the
SETUP/CAL
menu is
displayed. choose
DISPLAY
by pressing the number “
1
” key
+ ENTER
. The
FILTER
option can
be reached by scrolling down once to view the selection number and then pressing the “
4
” key
+
ENTER
. A conversion rate of 4, 15, 30 or 100 Hertz may be selected by entering its menu number
+ ENTER
. Enter the number of the desired A/D conversion rate and press
ENTER
. The normal
display will return. Changes may be noticed in the flickering of the LSB on the display. In most
situations, the flickering digits are an insignificant fraction of the display range and can be ignored.
Should the flickering become annoying or make reading the display difficult, choose a slower rate
of conversion.
10.2.6.
Front Panel Lock Out
Front Panel Lockout
is available only through serial communication. See the SERIAL
COMMUNICATIONS section of this manual and Appendix A.
10.3.
TOTALIZER SETUP
The
TOTALIZER
can be set to count up to a set point, to count down from a set point or to count up
continuously to the maximum ability of the display (999999). In order for the
TOTALIZER
to work, a
unit of measure must be chosen that includes a time element. For example, standard liters per hour
(SLH) will totalize; percent (%) will not. Each mode is described in the following text and can be followed
in the diagram below.
10.3.1.
Count Up to a Set Point
The
TOTALIZER
will be set to count from zero (0) to a set point. When the set point is reached, a
memory flag will be set to its logic high state.
This flag is readable only through serial
communications.
The
TOTALIZER
will continue to count up to the maximum ability of the display.
While the
SETUP/CAL
menu is displayed, choose
TOTAL
by pressing the number “
2”
key followed by
ENTER
. A
TOTALIZER
menu is displayed. Press the number “
1”
key to choose the
CNT UP
option.
The
COUNT UP/DN
window is displayed for the selected channel. Enter the desired set point using the
numbered key pads and press
ENTER
. The unit returns the previously programmed display with the
selected channel now displaying the Riemann Sum of the received transducer signal.
CAUTION
: While the front panel is locked out, all command functions, excluding
the override functions, are disabled. Commands can be overridden OPEN or
overridden CLOSED during front panel lockout. This was determined to be useful
both as a trouble shooting and as a safety tool. All other functions will be available
only through the use of serial communication.
NOTE:
Slower rates of conversion may affect the accuracy of the TOTALIZER in
situations where flow is changing often. Steady flow rates will see little loss of
TOTALIZER accuracy.