Page 40 of 42
Congratulations!
You have just calibrated your
Power
Pod
-400
to read zero at zero volts and the transducer’s max
display value at the transducer’s max signal input.
The only thing left to do is to insure that your transducer is calibrated and that it is wired correctly
to the
Power
Pod
-400
.
*
Individual channels may be calibrated with calibrated transducers connected, breakout
connectors (not supplied) and precision multi-meters with NIST traceable calibrations (also not
supplied). Such procedures are not covered in this document.
!
If the calibration for the channel just zeroed had been severely compromised prior to beginning
this procedure, zero may not be immediately displayed and you may have to repeat the procedure
after resetting the span. Check that the span has been set correctly!
#
If a signal level within .1% of the maximum DC voltage level is not attainable, the power supply
may need to be repaired.