134
ProEM System Manual
Issue 3
8.16 WinX/32 or LightField Crashes When Adding GigE Camera to
System
If you have a dual port Intel Pro/1000 Ethernet card and WinX/32 or LightField crashes
when trying to add a GigE camera to the system, you will need to modify the default IP
address for one of the ports.
To Modify the Default IP Address for a Port:
1. Click on the Windows Start button and click on Control Panel. See
Figure 8-11: Typical Windows Start button
2. On the Control Panel, click on Device Manager.
3. After the Device Manager opens, scroll down until PRO/1000 Grabber Devices (or
similar label) is displayed. See
.
Figure 8-12: Typical Device Manager Dialog
4
4
11-
0126
_00
8
9
4
4
11-
0126
_0
09
0
Summary of Contents for ProEM Series
Page 1: ...www princetoninstruments com ProEM System Manual 4411 0126 Issue 3 September 26 2019 ...
Page 10: ...10 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 16: ...16 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 30: ...30 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 50: ...50 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 88: ...88 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 114: ...114 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 122: ...122 ProEM System Manual Issue 3 Figure 7 8 LightField Settings 4411 0126_0078 ...
Page 136: ...136 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 146: ...146 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 152: ...152 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 161: ...This page is intentionally blank ...