56
ProEM System Manual
Issue 3
5.1.2
Spectroscopy Applications
The following paragraphs provide step-by-step instructions for operating ProEM in a
WinX spectroscopy setup for the first time. The intent of this simple procedure is to help
you gain basic familiarity with the operation of your system and to show that it is
functioning properly. Once basic familiarity has been established, then operation with
other operating configurations, ones with more complex timing modes, can be
performed. An underlying assumption for the procedure is that the camera is to be
operated with a spectrograph (such as a Teledyne Acton Research Series 2300
spectrograph) on which it has been properly installed. Refer to
, on page 147 for instructions for mounting a spectrograph
adapter to the ProEM. A suitable light source, such as a mercury pen-ray lamp, should
be mounted in front of the entrance slit of the spectrograph. Any light source with line
output can be used. Standard fluorescent overhead lamps have good calibration lines
as well. If there are no "line" sources available, it is possible to use a broadband source
such as tungsten for the alignment. If this is the case, use a wavelength setting of 0.0
nm for alignment purposes.
CAUTION!
!
Overexposure Protection:
Cameras that are exposed to
room light or other continuous light sources will quickly
become saturated. If the camera is mounted to a
spectrograph, close the entrance slit of the spectrograph to
reduce the incident light.
Assumptions
The following procedure assumes that:
•
You have already set up your system in accordance with the instructions in
previous chapters. This includes mounting the spectrograph adapter to the
camera. Refer to
Appendix C, Spectrograph Adapters
, on page 147.
•
You have read the previous sections of this chapter.
•
You are familiar with the application software.
•
The system is being operated in spectroscopy mode.
Getting Started
1. Set the spectrograph entrance slit width to minimum (10
m if possible).
2. Mount the camera to the spectrograph exit port.
3. Power ON the camera (the power switch is on the back of the power supply).
4. Turn on the computer power.
5. Start the application software.
Summary of Contents for ProEM Series
Page 1: ...www princetoninstruments com ProEM System Manual 4411 0126 Issue 3 September 26 2019 ...
Page 10: ...10 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 16: ...16 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 30: ...30 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 50: ...50 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 88: ...88 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 114: ...114 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 122: ...122 ProEM System Manual Issue 3 Figure 7 8 LightField Settings 4411 0126_0078 ...
Page 136: ...136 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 146: ...146 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 152: ...152 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 161: ...This page is intentionally blank ...