ProEM System Manual
Issue 3
•
Detector Temperature (Setup
►
Detector Temperature…)
The default temperature setting is read from the camera. When the array
temperature reaches the set temperature, the Detector Temperature dialog will
report that the temperature is LOCKED. Note that some overshoot may occur.
This could cause temperature lock to be briefly lost and then quickly
re-established. If you are reading the actual temperature reported by the
application software, there may be a small difference between the set and
reported temperature when lock is established. This is normal and does not
indicate a system malfunction. Once lock is established, the temperature will be
stable to within ±0.05°C.
See
Figure 5-6: Typical WinX/32 Detector Temperature Dialog
NOTE:
The Detector Temperature dialog will not display
temperature information while you are acquiring data.
•
Experiment Setup
►
Main Tab (Acquisition
►
Experiment Setup…)
— Exposure Time: 50 ms
— Accumulations & Number of Images: 1
•
Experiment Setup
►
ROI Tab (Acquisition
►
Experiment Setup…)
Use this function to define the region of interest (ROI).
— Spectroscopy Mode: Selected
— Clicking on Full loads the full size of the chip into the edit boxes.
•
Experiment Setup
►
Timing Tab (Acquisition
►
Experiment Setup…)
— Timing Mode: Free Run
— Shutter Control: Disabled Opened.
— Safe Mode vs. Fast Mode: Fast Mode
See
4
4
11-
012
6
_0
028
P
RO
EM:512B
P
RO
EM:1024B
Summary of Contents for ProEM Series
Page 1: ...www princetoninstruments com ProEM System Manual 4411 0126 Issue 3 September 26 2019 ...
Page 10: ...10 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 16: ...16 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 30: ...30 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 50: ...50 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 88: ...88 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 114: ...114 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 122: ...122 ProEM System Manual Issue 3 Figure 7 8 LightField Settings 4411 0126_0078 ...
Page 136: ...136 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 146: ...146 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 152: ...152 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 161: ...This page is intentionally blank ...