October 2017
RiverPro and RioPro ADCP Guide
Page 14
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
Changing the Baud Rate in the ADCPs
The RiverPro/RioPro can be set to communicate at baud rates from 1200 to 115200. The factory default
baud rate is always 115200 baud. The baud rate is controlled via the
CB-command
. The following proce-
dure explains how to set the baud rate and save it in the RiverPro/RioPro. This procedure assumes using
the program BBTalk that is supplied by Teledyne RD Instruments.
RiverPro
Teledyne RD Instruments (c) 2015
All rights reserved.
Firmware Version: 56.xx
>
>cr1
[Parameters set to FACTORY defaults]
Connect the RiverPro/RioPro to the computer and ap-
ply power.
Start the
BBTalk
program and
tions with the RiverPro/RioPro
Wake up the RiverPro/RioPro by sending a break signal
with the
End
key.
At the ">" prompt in the communication window, type
CR1
then press the Enter key. This will set the
RiverPro/RioPro to the factory default settings.
BAUD RATE
CB-command
1200
CB111
2400
CB211
4800
CB311
9600
CB411
19200
CB511
38400
CB611
57600
CB711
115200
CB811 (Default)
Send the CB-command that selects the baud rate you
wish. The table on the left shows the CB-command set-
tings for different baud rates with no parity and 1 stop
bit.
For example, to change the baud rate to 115200, at
the ">" prompt in the communication window, type
cb811
then press the Enter key.
The
CB?
command will identify the communica-
tion setting.
>cb?
CB = 411 ----------------- Serial Port Control (Baud
[4=9600]; Par; Stop)
>cb811
>CK
[Parameters saved as USER defaults]
>cb?
CB = 811 ----------------- Serial Port Control (Baud
[8=115200]; Par; Stop)
>
BBTalk
will send the command
CK
to save the new
baud rate setting.
Exit
BBTalk
.
The RiverPro/RioPro is now set for the new baud rate.
The baud rate will stay at this setting until changed
again with the CB command.
Exit
BBTalk
so the communication port is availa-
ble for use with other programs.