REMOTE CONTROL
79
[:SOURce]:CURRent:SLEWrate:RISing
Set
Query
Description
Sets or queries the rising current slew rate. This is
only applicable for CC slew rate priority mode.
Syntax
Query Syntax
[:SOURce]:CURRent:SLEWrate:RISing
{<NR2>(A)|MINimum|MAXimum}
[:SOURce]:CURRent:SLEWrate:RISing?
Parameter
<NR2>
Per step is between 0.001A/msec and
rated current divided by 100 msec.
MIN
Minimum rising current slew rate is
0.001A/msec.
MAX
Maximum rising current slew rate is
rated current divided by 100msec.
Return parameter <NR2>
Returns the step current in amps.
Example
SOUR:CURR:SLEW:RIS?
0.950
Sets the rising current slew rate to 0.950 A/ms.
[:SOURce]:CURRent:SLEWrate:FALLing
Set
Query
Description
Sets or queries the falling current slew rate. This is
only applicable for CC slew rate priority mode.
Syntax
[:SOURce]:CURRent:SLEWrate:FALLing
{<NR2>(A)|MINimum|MAXimum}
Query Syntax
[:SOURce]:CURRent:SLEWrate:FALLing?
Parameter
<NR2>
Per step is between 0.001A/msec and
rated current divided by 100 msec.
MIN
Minimum falling current slew rate is
0.001A/msec.
MAX
Maximum falling current slew rate is
rated current divided by 100msec.
Return Parameter <NR2>
Returns the step current