T3PS Series Programming Manual
84
[:SOURce]:VOLTage:SLEWrate:RISing
Set
Query
Description
Sets or queries the rising voltage slew rate. This is
only applicable for CV slew rate priority mode.
Syntax
Query Syntax
[:SOURce]:VOLTage:SLEWrate:RISing
{<NR2>(V)|MINimum|MAXimum}
[:SOURce]:VOLTage:SLEWrate:RISing?
Parameter
<NR2>
Per step is between 0.001V/msec and
rated voltage divided by 100msec.
MIN
Minimum rising voltage slew rate is
0.001V/msec.
MAX
Maximum rising voltage slew rate is
rated voltage divided by 100msec.
Return parameter <NR2>
Returns the slew rate in V/msec.
Example
SOUR:VOLT:SLEW:RIS MAX
Sets the rising voltage slew rate to its maximum.
[:SOURce]:VOLTage:SLEWrate:FALLing
Set
Query
Description
Sets or queries the falling voltage slew rate. This is
only applicable for CV slew rate priority mode.
Syntax
[:SOURce]:VOLTage:SLEWrate:FALLing
{<NR2>(V)|MINimum|MAXimum}
Query Syntax
[:SOURce]:VOLTage:SLEWrate:FALLing?
Parameter
<NR2>
Per step is between 0.001V/msec and
rated voltage divided by 100msec.
MIN
Minimum falling voltage slew rate is
0.001V/msec.
MAX
Maximum falling voltage slew rate is
rated voltage divided by 100msec.
Return parameter <NR2>
Returns the voltage slew rate in V/msec