GETTING STARTED
27
Load Connection
Recommended
Cables
Model
Specification Usage
GTL-104A 10A
Front panel terminal
GTL-105A
3A
Sense (for T3PS13206P
only)
Front panel
wiring
Use the GTL-104A cables for the
front panel source connections.
MASTER
0 - 32V , 3A
CH1
USB Type A only
(Greater than 4A)
3A MAX
Use the GTL-105A cables for the
sense connections.
S
S
Wire Gauge
Load wires must have enough current capacity to
minimize cable loss and load line impedance.
Voltage drop across a wire should not excess 0.5V.
The following list is the wire current rating at
450A/cm2.
Wire Size(AWG)
Maximum Current (A)
20
2.5
18
4
16
6
14
10
12
16