THCD-401 Instruction Manual
Page 44 of 47
Input channel range
uir
uir?
1-4,#
P1: Input channel number
P2: Real value for range to set
Query returns channel range for all input
channels
Input channel full scale
uif
uif?
1-4,#
P1: Input channel number
P2: Real value for full scale to set.
Query returns channel full scale for all input
channels
User input rezero function irz
irz?
1-4[,0]
P1: Input channel number
P2 (optional): clears current user input rezero
setting
Adaptive filter band
flb
flb?
#
P1: Band % (0.01-1.00), OFF, ON
Adaptive filter size
fls
fls?
0-6
P1: Buffer size in seconds
Relay switch threshold
rlt
rlt?
#
P1: Real number for switching point
Relay source
rls
rls?
1-4
P1: Input channel number
Relay hysteresis
rlh
rlh?
0.0-10.0
P1: Hysteresis percentage value
Ethernet IP address
eip
eip?
IP octets
P1: xxx.xxx.xxx.xxx
Ethernet subnet mask
esm
esm?
IP octets
P1: xxx.xxx.xxx.xxx
Retrieve all settings
ras
---
---
Comma separated string returned
See Table 3 for actual settings
Date of last calibration
---
dlc?
Returned date format: yymmdd