Appendix A: Errors
Vision G2 Testers User Guide—74-108-800 Rev. H—12 June 2019
138 of 158
Error
Code
Error
Category Title
Message
Fault
Check the Following
0034
3
Illegal Parameter at Event
X
Event X has illegal parameter value. Update
the protocol and use a valid parameter.
One of the parameters of the specified event is
incorrect.
Event and User Guide for proper syntax
0035
1
Illegal Parameter
The parameter value specified is invalid. Use
a valid parameter.
One of the parameters of the specified event is
incorrect.
Event and User Guide for proper syntax
0036
1
No Pre- or Post-Run
Samples
Pre-run or Post-run time sampling not
supported. Set sample event during test.
User tried to sample in pre-run or post-run.
Sampling events occur during testing.
0037
1
1 Pre/Post-Run Event at
a Time
Pre- and Post-run events can only be made
one at a time. Change Number of Events to 1.
The occurrence was greater than 1 when creating a
pre- or post-run event.
The occurrence field when creating an event
0038
4
Real Time Clock Error
Set the date and time of the instrument. If the
problem persists, contact Tech Support.
Time or date not set or corrupted, reported when the
instrument is powered on.
Time or date not set or corrupted, reported right
before the test run.
1. Time and date set
2. Instrument powered on for 5 minutes
3. Main board failure
0039
1
Incorrect Alarm
Date/Time
The alarm cannot be set in the past. Check
current time.
User attempted to set alarm in the past.
Alarm set for the future
0040
1
System Date Not Set
System date needs to be set before setting
alarm date.
System date has not been set.
System date
0041
1
System Date Not Set
System date needs to be set before setting
alarm time.
System date has not been set.
System date
0042
1
System Date Not Set
System date needs to be set before setting
start date.
System date has not been set or has been
corrupted.
System date
0043
1
System Date Not Set
System date needs to be set before setting
start time.
System date has not been set or hasbeen
corrupted.
System date
0044
1
Password Change Failed
Re-entered password didn't match the new
password.
The re-entered password didn't match the first
password entered.
Password and verification should match.
0045
1
Collect Must Be > Detect
Vol
The collect volume can't be lower than the
detect volume. Increase Collect volume or
lower Detect volume.
The detect volume is the same as or greater than
the collect volume.
Collect volume is greater than the detect volume.
0046
3
Call X Late X min X s
Call X was late X min X s due to a timing
conflict. Check event times.
The Call event occurred later than programmed
because another event was taking place.
Time between events
0047
1
Possible Timing Conflict
The event at time X might cause a timing
conflict.
One or more of the events may overlap causing a
delay.
Time between events
0048
3
Sample X Late X min X s
Sample X was late X min X s due to a timing
conflict. Check event times.
The sample collection was delayed because one or
more events were still being executed.
Time between events
0049
1
No Pre- or Post-Run
events
Pre-run or Post-run time events not supported
in events libraries.
Pre run or post run events can only be executed at
the protocol level.
Review user guide.
0050
1
Call Event Not Supported
Calling an event library from an event library is
not supported.
Call events can only be used in protocols
Review user guide.
0051
1
Event Library Name Too
Long
Event Library name cannot be more than 11
characters long.
Event library name is too long. Shorten the event
library name.
Review user guide.
0052
1
Cannot Load the Library
Library X could not be loaded. Check that it
exists.
The event library may have been deleted after the
protocol was created.
Review existing protocols and event libraries.
0053
1
No Pre- or Post-Run
Samples
Library X has a sample event. It cannot be
called at Pre-run or Post-run.
Sample events are not permitted in pre or post run.
This includes event libraries called in the pre or post
run.
Review user guide.
Summary of Contents for VISION G2 CLASSIC 6
Page 1: ...VISION G2 CLASSIC 6 AND ELITE 8 DISSOLUTION TESTERS USER GUIDE 74 108 800 Rev H 12 June 2019...
Page 148: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 147 of 158...
Page 149: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 148 of 158...
Page 150: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 149 of 158...
Page 151: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 150 of 158...
Page 152: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 151 of 158...
Page 153: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 152 of 158...
Page 154: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 153 of 158...