Operation
Vision G2 Testers User Guide—74-108-800 Rev. H—12 June 2019
70 of 158
Tools
Tools button
Touching the Tools button on the main menu displays the Tools
screen.
This provides access to the Magazine Calibration screen first
(only available when an AutoMag or SuperMag is installed and
Auto-Probes are configured in the Config menu). This screen
allows the user to calibrate the AutoMag or SuperMag and
provides access to basic control for movement of the magazine.
Magazine Calibration screen
The Magazine Calibration screen is also accessed by touching
the Cal Mag button at the bottom of the screen.
Mag Sample Travel: This field is the depth at which the
Magazine will travel to when the sample button is touched.
Up: Moves the Magazine up.
ADD: Moves the Magazine down to the height to trigger Auto
Dosage Delivery covers.
Home: Homes the Magazine by raising it up against the drive
head them moving it down a set amount.
Sample: Moves the Magazine to the sample position specified
in the Mag Sample Travel field.
Summary of Contents for VISION G2 CLASSIC 6
Page 1: ...VISION G2 CLASSIC 6 AND ELITE 8 DISSOLUTION TESTERS USER GUIDE 74 108 800 Rev H 12 June 2019...
Page 148: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 147 of 158...
Page 149: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 148 of 158...
Page 150: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 149 of 158...
Page 151: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 150 of 158...
Page 152: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 151 of 158...
Page 153: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 152 of 158...
Page 154: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 153 of 158...