Operation
Vision G2 Testers User Guide—74-108-800 Rev. H—12 June 2019
71 of 158
Calibrate: This button runs the magazine calibration procedure.
The AutoMag is calibrated before it leaves the factory. However,
if it needs to be recalibrated, use the following procedure from
the Magazine Calibration screen:
1.
Place the AutoMag calibration tool in the center of
the vessel plate. (See Fig. D-1 for placement.)
NOTE: Before touching the Calibrate button, ensure
that the magazine calibration tool is in place on the
vessel plate below the drive screw.
Fig. D-1 AutoMag Calibration Tool in Place