Operation
Vision G2 Testers User Guide—74-108-800 Rev. H—12 June 2019
87 of 158
Infinity Test Length: The length of the infinity test which follows
the dissolution test. If no infinity test is desired set this value to
0:00. The range is 0:00 to 9:59 in h:mm format.
NOTE: The infinity test begins after the dissolution test is over.
For example, if a dissolution test was 2 hours, with a 30-minute
infinity test, the user would need to set a sample point at 2:30
(h:mm) to collect an infinity test sample.
Infinity Test Speed: Sets the speed of the infinity test that follows
the normal test. The range is 0, 25 to 250 rpm.