Operation
Vision G2 Testers User Guide—74-108-800 Rev. H—12 June 2019
91 of 158
(T) Bath Temp: The parameter is a temperature value. It allows
the user to set a new dissolution tester bath temperature.
(F) Change Filter: Instructs the Vision AutoFilter to change a
filter. The parameters are the testers (A, B, or C) for which the
filter is changed.
For example, if B was entered, the AutoFilter would change a
filter for the sample on tester B, but would not change the filter
for tester A.
(S) Collect Volume: This allows the user to set a new collection
volume for the test.
(B) Delay: This allows the user to set a delay for the start of the
test. The delay waits for the specified amount of time. Range 1
to 12,000 sec. It is valid for Pre-Run only.
NOTE: To stop paddle rotation temporarily during a dissolution
test, use the Pause event.
(S) Detect Volume: The parameter is the detection volume. It
allows the user to set a new Detect Volume for the test.
(S) Dilute: The options are (dilution) Volume, Rinse (volume),
Source, StartRow, and RowCount. It allows the user to perform
a dilution at the time of the event rather than after each sample
(dilution event). The range is 0 to 99.9 mL.
NOTE: The Dilute event is only compatible with port B or port C
set as a media source. It will not function with Home Port 2 set
as a media source.
For example, if the user programmed a Dilute event at 2 hours
with 5, 7, C, 3, 10, the AutoPlus would first rinse from C to D
with 7 mL, then pump 5 mL from source C in to rows 3 through
13.
(S) Dilution: The options are Volume, Rinse, and Source. It
allows the user to perform a dilution during the test only after
each sample. The range is 0 to 99.9 mL. Set volumes to 0.0 to
turn off.
For example, for a dilution volume of 5 mL, a rinse of 7 mL, and
a source C, the parameter would read: 5,7,C.
(F) Discard Filter: The AutoFilter discards the current filter. The
parameters are the tester.
For example, if an operator entered A, the filter would be
discarded for tester A sample points.
Summary of Contents for VISION G2 CLASSIC 6
Page 1: ...VISION G2 CLASSIC 6 AND ELITE 8 DISSOLUTION TESTERS USER GUIDE 74 108 800 Rev H 12 June 2019...
Page 148: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 147 of 158...
Page 149: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 148 of 158...
Page 150: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 149 of 158...
Page 151: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 150 of 158...
Page 152: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 151 of 158...
Page 153: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 152 of 158...
Page 154: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 153 of 158...