Description
The B2A Portable is a thermocouple vacuum sensor controller with a range of 1
×
10
−
3
Torr to 20 Torr.
It utilizes a Televac thermocouple vacuum sensor, part number 2-2100-10 or 2-2126-0YY. It has an
easy-to-read analog display that indicates pressure in units of mTorr.
1.3
Operating Specifications
Table 1: Operating Specifications
Operating Range
1
×
10
−
3
to 20 Torr
Communications
N/A
Programmable Set Points
N/A
Analog Output
N/A
Supply Voltage for Charging
90 V to 135 V AC, 60 Hz
Calibration Medium
Dry air or nitrogen
Overpressure
Sensor-dependent
Battery Life
12 hours
Charging Time
15 hours
Operating Temperature
0
◦
C to 50
◦
C
Storage Temperature
−
20
◦
C to 60
◦
C
Bakeout Temperature
Sensor-dependent
Response Time
≤
1 second
Accuracy
1 mTorr to 10 mTorr
±
1 mTorr
10 mTorr to 500 mTorr
±
10%
500 mTorr to 1 Torr
±
20%
Display Readable Distance
1 m (3 feet)