Page 14 of 40
Y
TELEVA
Crystal gauge zero adjustment mode In this mode, it is possible to manually make a zero adjustment of the
crystal gauge. Usually this adjustment is not necessary, because it is performed automatically every time the
pressure reaches 3.0 x 10-3 Torr. When the crystal sensor is replaced for maintenance or if the zero has
shifted, a manual zero adjustment might be required.
Crystal gauge Zero adjustment mode: When
“parameter lock” is enabled this adjustment is not possible.
Make sure if lock is disabled before the programming. (Refer to 4.2)
The character
“02 on” on the display shows the unit is in this mode. Pushing [ENT] key triggers the
zero adjustment instructing the sensor to fit the measurement to zero at the event.
Note: The same adjustment can be triggered from I/O pin #6. See section 8.
Analog output setting mode. The analog output can be selected from the Log output mode or the combined
output mode. When the Log output is selected, it is necessary to set the width voltage of one decade and
the full scale voltage. If the optional display unit DS-10 is used, the output mode will have to be set
accordingly.
In this section only the operation will be explained. A detail of the analog output is described in the next
section.
The analog output setting mode: When
“parameter lock” is enabled any parameter change can not be done
and only displaying the current parameters is possible. Make sure if lock is disabled before the programming.
(Refer to 4.2)
HV
SP2
SP1
SP3
Torr
X10
FUNC
ENT
Select Output Mode
Lit Blink
8-7
When the analog output setting mode is selected,
“03” is appear on the left
side two digit of the display. The right side two digit of the display blink and
show the out put mode.
01: Log output 02: Combined output 03: Output for the display unit DS-10.
The change of the setting should be made according to the following
procedures.