Set Full Scale
Set 1 decade
voltage
Set Full Scale
Set 1 decade
voltage
SP3
SP2
SP1
HV
Torr
X10
FUNC
ENT
Select
Combined Mode
For Output
Lit Blink
Lit Blink
Lit Blink
Lit Blink
Lit Blink
To select the Log output mode and set the width of one decade
0.5(V)
(Width of one decade can be set either 0.5V or 1.0V.)
(1) Set the blinking display to 01 by [
↑] key then push [ENT] key to
select Log output mode.
(2) Blinking value becomes to show the one decade width. Set the value
0.5(V) then push [ENT] key.
3) Blinking value becomes to show the full scale voltage. Set the value
between 7 to 10V then push [ENT] key. The unit returns to measure
mode.
To select the Log output mode and set the width of one decade
1.0(V)
(1) Set the blinking display to 01 by [
↑] key then push [ENT] key to
select Log output mode.
(2) Blinking value becomes to show the one decade width. Set the value
1.0(V) then push [ENT] key.
(3) Blinking value becomes to show the full scale pressure. When
1.0V decade is selected maximum voltage is automatically set to 10V
and full scale pressure corresponding to 10V can be set here. Select
value from 00(1.0x10
0
Torr) to 03(1.0x10
3
Torr). Push [ENT] key to
register the value after selecting the full scale pressure. The unit returns
to measure mode.
To select the combined output mode set the blinking digit at 02 by
[
↑] key then push [ENT] key to register the mode. The unit returns to
measure mode.