INSTRUMENT MANUAL
VACUUM GAUGE MODEL MM200
160Phone:(215) 947-2500 fax:(215) 947-7464 e-mail:sales@televac.com web site:www.televac.com
MM-200_im REV M
Page 103 of 160
TELEVAC
RS232
The MM200 uses a DE9S connector (female) for RS232
HARDWARE
and the RS485 hardware interface. The pinouts are in
accordance with Fig. 6.1 and Table 6.1.
PIN
FUNCTION
2
Tx - from MM200
3
Rx - to MM200
4
DTR - to MM200
5
GROUND
6
DSR - TO MM200
7
DO NOT USE
8
DO NOT USE
1
RS485 (Same as "B")
9
RS485 (Same as "A") *
5 1
o o o o o
o o o o
9 6
* For PCB # 3900G105 (See copper side)
- Not connected for "F" revision and lower.
RS485
The MM200 uses a 3 pin connector for the RS485
HARDWARE
hardware interface. The pinouts are in accordance with Figure and table. One
possible way to interconnect is shown below using a 485TBLED converter.
PIN
FUNCTION
B
Lead "B"
A
Lead "A" (Less + at IDLE mode)
G
Ground