INSTRUMENT MANUAL
VACUUM GAUGE MODEL MM200
160Phone:(215) 947-2500 fax:(215) 947-7464 e-mail:sales@televac.com web site:www.televac.com
MM-200_im REV M
Page 107 of 160
TELEVAC
R2AC<cr>
0-10V
of 2As to be MP2C traditional.
R2AL<cr>
0-10V
of 2As to be logarithmic.
R2AN<cr>
0-10V
of 2As to be linear to (1) Torr.
R2AT<cr>
0-10V
of 2As to be 2A traditional.
R3DL<cr>
0-10V
of 3D to be logarithmic.
R3DM<cr>
0-10V
of 3D to be multi linear.
R4AL<cr>
0-10V
of 4As to be logarithmic.
R4AN<cr>
0-10V
of 4As to be standard.
R7BE<cr>
0-10V
of 7Bs to be extended log.
R7BL<cr>
0-10V
of 7Bs to be logarithmic.
R7BM<cr>
0-10V
of 7Bs to be multi linear.
R7BN<cr>
0-10V
of 7bs to be linear.
R7EL<cr>
0-10V
of 7Es to be logarithmic.
R7EM<cr>
0-10V
of 7Es to be multi-linear.
SET POINTS
PCA<cr>
Control all relays via RS232.
PCn<cr>
Control relay "n" via RS232.
PFn<cr>
Turn off controlled relay "n" via RS232.
PNn<cr>
Turn on controlled relay "n" via RS232.
PUA<cr>
Remove all relays from RS232 control.
PUn
Remove relay "n" from RS232 control.
SAxSy
Assign relay x to station y.
SPx<cr>
Send the station assignment of relay x.
SPxN<cr>
Send the "ON" setting of relay x.
SPxF<cr>
Send the "OFF" setting of relay x.
SSxFyyyyu<cr>
Set relay #x "OFF" to yyyy, units H or L.
SSxNyyyyu<cr>
Set relay #x "ON" to yyyy, units H or L.