INSTRUMENT MANUAL
VACUUM GAUGE MODEL MM200
160Phone:(215) 947-2500 fax:(215) 947-7464 e-mail:sales@televac.com web site:www.televac.com
MM-200_im REV M
Page 118 of 160
TELEVAC
"SPx<cr>"
8. Send the station assignment of a particular station. This causes the modular unit
to respond with the station number assigned to relay "x" followed by a "<cr>".
"SPxN<cr>"
9. Send the ON setting of a particular relay. This causes the modular unit to
respond with the ON set point setting of relay "x". The response to this command
is a four-digit number including lead zeros followed by a "<cr>". The format of the
response is the same as for the "SSxFyyyyu" command below.
"SPxF<cr>"
10. Send the OFF setting of a particular relay. This causes the modular unit to
respond with the OFF set point setting of relay "x". The response to this command
is a four-digit number followed by a "<cr>". The format is the same as for the
SSxFyyyyu" command below.
"SSxFyyyyu<cr>"
11. Set the OFF set point of a particular relay. Where "x" is the relay number for
which it is desired to set the dropout (de-energize) point, and "nnnn" is the setting.
The response to this command is "A<cr>". See the chart below (next page) for
ranges of settings for various sensors.
"SSxNyyyyu<cr>"
12. Set the ON set point of a particular relay. Where "x" is the relay number, and
"nnnn" is the setting. See the chart below (next page) for ranges of settings for
various sensors.
SET POINT SETTING VALUES
SENSOR
UNITS
RANGE
SETTING
2A
MICRONS
0-1000
0001L to 0999L
TORR
1.0-20.0
0010H to 0200H
3D
TORR
0.0e-11 to 9.9e-2
0.0-B to 9.9-2
4A
MICRONS
1-1000
0001L to 0999L
TORR
1-1000
0001H to 0999H
1E
TORR
1-1000
0001L to 0999L
7B
TORR
1e-6 to 1.5e-3
0001L to 0990L
7E
TORR
1.0e-8 to 9.9e-6
0001L to 0990L
TORR
1.0e-5 to 9.9e-3 0001H to 0990H