900 TROUBLESHOOTING
901 PRELIMINARY TROUBLESHOOTING
Refer to prior sections of the manual for preliminary troubleshooting help in accordance with Table 9.1.
Table 9.1
200
Initial checkout
300
Installation
400
Operation
500
Programming
600
Communications
700
Calibration
800
Specifications
Next are some possible problems and the suggested solutions.
P -
All displays are blank.
S -
Check power switch, power source and power cable.
P -
Ion gauge reads "OFF" and controlling TC reads 20.0 Torr and is flashing.
S -
Normal. Pressure is above the ion-on setting of 10 microns and the
controlling (chamber) TC is above 20.0 Torr.
P -
Ion gauge reads "OFF" and Ion-on source TC reads above 10 microns.
S -
Normal. Pressure is above the ion-on setting of 10 microns.
P - Display shows readings other than pressure.
S - Press TEST/OP. If MICRONS PER HOUR is lighted, press LEAK RT
pushbutton.