INSTRUMENT MANUAL
VACUUM GAUGE MODEL MM200
160Phone:(215) 947-2500 fax:(215) 947-7464 e-mail:sales@televac.com web site:www.televac.com
MM-200_im REV M
Page 18 of 160
TELEVAC
300 INSTALLATION
301 MODULE
CONFIGURATION
A controller is normally ordered with various optional sensor interface modules and they are factory installed before
shipment. Before installation, verify that the selection of modules will accomplish the desired results in your
application. If additional modules are required or you have any questions about using this gauge in your
application, please call Televac at 215-947-2500. If you wish to add modules, refer to Section 307 AND 308.
One possible configuration for this gauge was shown in Section 203. The allowable configurations are restricted by
the number of cold cathode and/or hot cathode ionization stations installed.
The MM200 has six slots, which can be populated with various Televac modules. There is also a half slot that can
be used to insert a communications module. This is the only slot into which the communications module can be
inserted. This board will currently accommodate an RS232/RS485 interface. The other slots can be fitted with
various combinations of the following currently available boards:
PCB #
TYPE 6200 STA SLOT
DESCRIPTION
1E -220 2 1 A dual diaphragm module.
1F -224 2 1 A dual diaphragm module.
2A -186 2 1 A dual thermocouple module.
3D -226 1 4 A single hot cathode module with Joule de-gassing.
3E -232 1 4 A single hot cathode module with EB de-gassing.
4A -215 2 1 A dual convection module.
7B -227 1 2 A single cold cathode module.
7E -223 1 2 A single cold cathode module.
7F -245 1 1 A single cold cathode module (7F UHV or 7E as req'd)
7F -253 1 1 An interface to the 7ER & 7FR remote modules.
SP
-211 4 1 A four relay Process Control relay board.
3F -290 2 2 A dual Mini BA module
The 3D and 3E modules occupy four slots each.
The 7B and 7E modules occupy one and a half slots. One of these can be combined with the RS232/RS485 digital
interface module so that no space is wasted. In other locations, they will take up two slots.
The 7F module takes up only one slot. It is used with the 7F UHV tube down to 10
-11
Torr or with multiple 7E gauge
tubes where space available in the MM200 is limited.
The 7FCS module has a phone jack labeled “starter” at the back panel to plug the igniter cable in.
The 7F 6200-253 module can interface with 7ER & 7FR gauge tubes which have the remote electronics
modules.