The maximum number of stations that can be configured is ten. This can be done
only if all stations are any mixture of dual modules (e.g., 1E, 1F, 2A, 4A, 5A/B).
WARNING:
If any cold cathode gauge modules are installed, no sensors should be
assigned to station number ten (10).
If there is one relay module installed, then five other slots are available, and still up
to ten stations can be installed. With two relay modules, four slots are available,
and up to eight stations can be accommodated.
Setpoints must be consecutively numbered starting with number 1 (1 through 4) or
number 5 (5 through 8).
WARNING:
All relays must be assigned to a station when first installed, even if they will
not be used (any station will suffice).
Allowable configurations for the MM200 are shown in
Table 3.1.