INSTRUMENT MANUAL
VACUUM GAUGE MODEL MM200
160Phone:(215) 947-2500 fax:(215) 947-7464 e-mail:sales@televac.com web site:www.televac.com
MM-200_im REV M
Page 33 of 160
TELEVAC
Table 3.4 Tube Type
Rotary Sensitivity
Sensitivity Glass Tube
Nude Tube
Switch Switch
0 10/12.5
12.5
Plain 2160-1xx
1 10/12.5
12.5
Coated 2162-1xx
2 10/12.5
10
Plain
2161-2xx I
2
R 2164-352
3 10/12.5
10
Coated 2163-2xx
0 20/25
25
Plain
EB 2166-452
1 20/25
25
Coated
2 20/25
20
Plain
3 20/25
20
Coated
Note: Switch position with power off or turn power ON/OFF after switching.
There is no filament selector switch for use with dual filament sensors on this unit as this is done through front
panel commands. It is necessary to switch the filaments when the filament power is off. It may also be necessary
to reverse the cable connector.
An analog output is available at the connector in the lower left of the panel. There are three terminals available.
The lower terminal is a common ground and negative terminal for both the upper two terminals. The upper two
terminals are for + 10 volts and + 10 millivolts respectively.
Refer to Section 502.3 Hot Cathode Setup for further information.
Fig. 3.09 Rear Panel of the Resident 3E Hot Cathode Gauge Module