406 INITIAL OPERATIONAL SETUP SEQUENCE
Turn power ON. Switch located on gauge controller rear panel. Observe pressure readings (with a cable and a
sensor plugged in).
PRESSURE READINGS
Before pumpdown, the chamber thermocouple gauge will read 20.0 Torr and the cold cathode gauge will read
"OFF". As pumpdown proceeds, the thermocouple gauge will start to read pressures below 20.0 Torr. When it gets
to 10 microns, the cold cathode gauge will be powered and will start to display a reading. Other gauge tube
readings are observed using the UP/DOWN pushbuttons.
RELAY SETUP
The SETUP pushbutton is used to control functions. They are controlled in the following sequence:
1. Cold Cathode gauge setup - if module is present
2. Hot Cathode gauge setup - if module is present
3. Relay setpoints/assignment - if module is present
If module is not present, this step is skipped. See section 502.3 for Hot Cathode and cold cathode setup
procedures if these modules are present. For relay setup proceed as follows:
1.
Push SETUP button as many times as is required to see the following display
a.
SP1
b.
STATION 1 (blinking) or any other station [Change assigned station, if
required.
Use UP/DOWN pushbuttons]
c.
Sensor Type (e.g., 2A)
2.
Push SETUP button to display
a.
"on" value of SP 1
b.
STATION 1 (not blinking)
c.
Change "on" value, as desired, with either set of UP/DOWN arrows.
3.
Push SETUP button to display
a.
"OFF" value of SP1
b.
STATION 1 (not blinking)
c.
Change "OFF" value, as desired, with UP/DOWN arrows.
4.
Push TEST/OP button to return to NORMAL display at any time.
5.
Push SETUP button to display SP 2, 3 & 4 and change values as desired as described in steps 3
& 4.
6.
Push SETUP button to return to NORMAL display.
NOTE:
Station #1 is chosen as an example and may be other than a thermocouple
station. Also, any active station may be selected to control the relays as noted in
1b. Additional information can be found in Section 502.3 under Relay Setpoints.