INSTRUMENT MANUAL
VACUUM GAUGE MODEL MM200
160Phone:(215) 947-2500 fax:(215) 947-7464 e-mail:sales@televac.com web site:www.televac.com
MM-200_im REV M
Page 98 of 160
TELEVAC
R2AR<cr>
Recorder output of 2As to be reversed linear to (1) Torr.
R2AT<cr>
Recorder output of 2As to be 2A traditional.
R3DL<cr>
Recorder output of 3D or 3E to be logarithmic.
R3DM<cr>
Recorder output of 3D or 3E to be multi linear.
R3Dx<cr>
Recorder output of 3D or 3E to be linear to 10-x Torr
where 8 => x >= 2 and 10V is at 10-x Torr.
R4AL<cr>
Recorder output of 4As to be logarithmic.
R4AN<cr>
Recorder output of 4As to be multi linear.
R4AT<cr>
Recorder output of 4As to be linear to 10 Torr.
R4AU<cr>
Recorder output of 4As to be linear to 1 Torr.
R5AL<cr>
Recorder output of 5As & 5Bs to be logarithmic.
R5AN<cr>
Recorder output of 5As & 5Bs to be linear.
R7BA<cr>
Recorder output of 7Bs to be 7AC traditional.
R7BE<cr>
Recorder output of 7Bs to be extended log.
R7BL<cr>
Recorder output of 7Bs to be logarithmic.
R7BM<cr>
Recorder output of 7Bs to be multi linear.
R7BN<cr>
Recorder output of 7Bs to be linear.
R7BT<cr>
Recorder output of 7Bs to be B7B2 traditional.
R7Ex<cr>
Recorder output of 7Es to be linear to 10-x Torr
where 7 => x => 2 and 10V is at 10-x Torr
R7FM<cr>
Recorder output of 7Fs to be multi linear.
1V/Decade and 9V is at 10-2 Torr.
R7FL<cr>
Recorder output of 7Fs to be logarithmic. 9V is at 10-2 Torr.
R7Fx <cr>
Recorder output of 7Fs to be linear to 10-x Torr
Where 8 => x >= 2 and 10V is at 10-x Torr.
R7EL<cr>
Recorder output of 7Es to be logarithmic.
R7EM<cr>
Recorder output of 7Es to be multi-linear.
SET POINTS
CPn<cr>
Clear setpoint values for relay "n".
PCA<cr>
Control all relays via RS232.
PCn<cr>
Control relay "n" via RS232.
PFn<cr>
Turn off controlled relay "n" via RS232.
PNn<cr>
Turn on controlled relay "n" via RS232.
PUA<cr>
Remove all relays from RS232 control.
PUn
Remove relay "n" from RS232 control.
SAxSy
Assign relay x to station y.
SPx<cr>
Send the station assignment of relay x.
SPxN<cr>
Send the "ON" setting of relay x.
SPxF<cr>
Send the "OFF" setting of relay x.
SSxFyyyyu<cr>
Set relay #x "OFF" to yyyy, units H or L.
SSxNyyyyu<cr>
Set relay #x "ON" to yyyy, units H or L.