background image

 

phone: (215) 947-2500        fax: (215) 947-7464       e-mail:sales@televac.com       web site: www.televac.com

  

8/15 

5

 

 

 

SENSOR OPERATION AND INSTALLATION 

 

 

1.0 

Connect the gauge to the vacuum system.

 

It must be mounted vertically with the stem side 

down

.

 

 

NOTE: DO NOT TURN INSTRUMENTATION WHEN MOUNTING ON A THREADED INSTALLATION 

TO INSTALL, USE A WRENCH ON THE FLATTED PORTION OF THE SENSOR TUBE

 

 

2.0 

Assure that there are no leaks. 

 

3.0 

Apply 24 DC volt power between pin # 3 (+) and pin # 4 (-) of the connector cable. 

 

4.0 

Configure output lines; pin # 5 (for linear 1 T or log 1000T output), pin # 8 (signal gnd) and pin # 7 

(for non linear output). The table of voltage versus pressure is shown in chapter # 5. The linear 

signal output is 10 volts at 1000 microns and decreases linearly to zero at zero microns. The non-

linear signal output is 10 volts at atmospheric pressure and decreases non-linearly to zero volts at 

zero  microns.  If  it  is  a  log  1000T  unit,  the  output  decreases  logarithmically  from  6.0  volts  at 

1000Torr to zero volts at zero microns. 

 

5.0 

Pump down the vacuum system while observing the LED display and the signal voltage reading 

on the output. (see chapter # 5 for output values) 

 

6.0 

While  the  pressure  is  above  1000  Torr  the  display  will  show  “HI”.  Below  1000  Torr  the  signal 

output  will  correspond  to  values  given  in  chapter  #  5,  depending  upon  chosen  output.  As  the 

pressure drops below 1 Torr the red “MICRONS” LED lights and the displayed readings now drop 

from 990 down to zero while the linear output voltage drops correspondingly from 9.90 Volts to 

zero. 

 

7.0 

The trip pressure for the setpoint relay can be displayed by pressing the “setpoint display” button 

on the front panel of the instrument or by reading the Set Point voltage at pin # 9 of the cable 

connector. The display will blink the reading until the button is pushed again or until it times out 

automatically after 60 seconds. If the Set Point adjustment screw is turned to change the setting 

the new value displayed will be retained in memory. The setpoint adjustment range is 0 to 1 Torr 

for  the  units  with  linear  recorder  output  and  0-1000  Torr  for  units  with  logarithmic  recorder 

outputs. The relay contact connections are as follows; to the common # 6,  # 1, normally open, 

and # 2, normally closed.  

 

 

8.0 

If the display shows “-1” or “LO” then the gauge is out of adjustment. With a known vacuum of at 

least 1 X 10-5 Torr, adjust the “Zero” adjustment screw on the front panel.

 

 

 

 

9.0 

With the sensor tube exposed to atmospheric pressure, the display reading should be near 760 

Torr (which may vary with geographic location). Adjustment for the correct reading is done with 

the front panel “ATM” settings. Make sure that the tube is vertical, with stem down. 

 

 

 

 

SEE NEXT PAGE FOR GAUGE DIAGRAMS

 

 

CHAPTER 4

                                      

TELEVAC

 

Summary of Contents for MP4AR Series

Page 1: ...INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range 1 x 10 3 Torr to 1 x 10 3 Torr TELEVAC A DIVISION OF THE FREDERICKS COMPANY 2400 PHILMONT AVE HUNTINGDONVALLEY PA 19006 ...

Page 2: ...TEM DESCRIPTION 1 1 4AR Remote Convection Electronics 2 1 Cable specified at ordering 3 1 Instruction Manual 4 1 Sensor 1 2 4 3 phone 215 947 2500 fax 215 947 7464 e mail sales televac com web site www televac com 8 15 ...

Page 3: ...NT OPERATION 2 02 SAFETY 3 03 PRINCIPLES OF SENSOR OPERATION 4 04 SENSOR OPERATION AND INSTALLATION 5 05 RECORDER OUTPUTS 7 06 MAINTENANCE 8 07 CALIBRATION 8 08 TROUBLESHOOTING 8 09 SPECIFICATIONS 9 10 DIMENSIONS 10 11 UNPACKING AND INSPECTION 11 12 WARRANTY INFORMATION 11 PLEASE READ THIS MANUAL THOROUGHLY BEFORE USING GAUGE REPORT ANY PROBLEMS IMMEDIATELY ...

Page 4: ...LC chart recorder or data acquisition system The MP4AR series of instruments use the time proven reliable and rugged TELEVAC 4A Sensor which can be changed in seconds These instruments operate from low voltage DC and are easily integrated into the largest process system with a minimal effort Likewise the MP4AR Remote Gauge can be used as a small stand alone vacuum measurement solution Features Wid...

Page 5: ...y standards of design manufacture and intended use of the instrument Televac disclaims all liability for the customer s failure to comply with these requirements READ Instructions Read all safety and operating instructions before operating the product RETAIN instructions Retain the Safety and Operating Instructions for future reference HEED warnings Adhere to all warnings on the product and in the...

Page 6: ...ate measurements Sensors are available in stainless steel or nickel plated brass and can withstand 150 psi overpressure Objects and Liquid Entry Never push objects of any kind into this product through openings as they may touch dangerous voltage points or short out parts that could result in a fire or electric shock Be careful not to spill liquid of any kind onto the products Do not substitute pa...

Page 7: ...rrespond to values given in chapter 5 depending upon chosen output As the pressure drops below 1 Torr the red MICRONS LED lights and the displayed readings now drop from 990 down to zero while the linear output voltage drops correspondingly from 9 90 Volts to zero 7 0 The trip pressure for the setpoint relay can be displayed by pressing the setpoint display button on the front panel of the instrum...

Page 8: ... Relay normally closed contact 3 24 volt power 4 Power common 5 Log Linear signal output 6 Relay common contact 7 Non linear signal output 8 Signal return 9 Setpoint output CHAPTER 4 cont TELEVAC Setpoint Display Potentiometer Setpoint Adjust Potentiometer LED display Zero Adjustment Potentiometer Pin s LED lights for microns Phillips Head screw Atmosphere Adjustment Potentiometer SET POINT RELAY ...

Page 9: ...on linear Voltage Logarithmic 1 x 10 3 01 0 5 x 10 3 06 70 2 x 10 2 24 1 3 4 x 10 2 45 1 6 6 x 10 2 64 1 78 8 x 10 2 82 1 9 1 98 2 2 1 67 2 3 5 2 99 2 7 8 3 84 2 9 2 5 45 3 3 10 7 32 4 30 7 84 4 48 50 7 93 4 7 80 8 01 4 9 120 8 08 5 08 160 8 19 5 2 180 8 24 5 26 300 8 76 5 48 500 9 57 5 7 760 10 5 88 1000 10 22 6 LINEAR Pressure microns Voltage 1 01 5 05 10 1 20 2 30 3 40 4 50 5 60 6 70 7 80 8 90 ...

Page 10: ...hould be required unless it has been changed For a reliable recalibration at a later date it is necessary to have a reference standard vacuum system whose pressure is known to be accurate For such a calibration the instrument sensor tube and cable should be returned to the factory for re certification However if the vacuum calibration verification must be done at the customer s facility a knowledg...

Page 11: ... Power 24VDC 100mA customer supplied Materials Stainless steel or nickel plated brass Set point Relay Form C contact Overpressure 24 v 5 a rating Indicator lamp provided 150 psi Accuracy Adjustment Potentiometers 1 micron 0 to 10 microns Zero Atmosphere Setpoint 10 of reading 10 microns to 10 Torr Response Time for a Pressure 20 of reading 10 Torr to 100 Torr Change from 10 5 Torr to Atmosphere 10...

Page 12: ...phone 215 947 2500 fax 215 947 7464 e mail sales televac com web site www televac com 8 15 10 Dimensions CHAPTER 10 TELEVAC ...

Page 13: ...authorized to give any guarantee or warranty or make any representation in addition or contrary to those stated herein other than those expressly stated herein there are no other warranties of any kind express or implied and specially excluded but not by way of limitation are the implied warranties of fitness for a particular purpose and merchantability It is understood and agreed the seller s lia...

Reviews: