background image

2

SETUP

2

Setup

2.1

Installation

Each MV2A is designed to be used in conjunction with a 2A sensor. The 2A sensor can be mounted
in any orientation. The sensor should be mounted close to the area where vacuum measurement is
desired. Each 2A sensor has a key that only allows the octal end of the MV2A cable to mate with the
sensor in the proper orientation. Rotate the octal end of the MV2A cable until the correct alignment is
obtained and the MV2A cable is able to slide onto the 2A sensor. Connect the RJ14 end of the MV2A
cable into the port on the back of the MV2A labeled "input". It is recommended that the ground con-
nector on the cable is also connected to the port on the back of the MV2A labeled with the universal
earth ground symbol.

Try to avoid connecting the 2A sensor to the vacuum chamber by long or narrow piping as this may
affect the accuracy and response time of the unit. Avoid mounting the unit near a heater within the
chamber as this may affect the measurement accuracy. Excessive vibration of the sensor may affect
accuracy and decrease the life of the sensor. Exposure to oils and other contaminants will decrease
the accuracy and decrease the life of the sensor. Do not expose the sensor to corrosive gases.

2.2

Electrical Information

The MV2A has several connectors located on the rear of the unit. A connection diagram is provided
for convenience.

Figure 2: MV2A Electrical Connection Information

Label

Description

1

AC Input

2

SP1 C

3

SP1 NC

4

SP1 NO

5

SP2 C

6

SP2 NC

7

SP2 NO

8

Analog  (0 to 5 V)

9

Analog Output – (0 to 5 V)

10

Earth Ground

11

Sensor Input

2.3

Set Points

The instrument is equipped with dual set points which are adjustable from the front panel. These two
set points are control relays, which can be found on the rear of the controller. These can be wired to
be normally open or normally closed.

To adjust the set point values, hold either

SP1

or

SP2

while toggling the

UP

or

DOWN

buttons

to the desired value on the screen.

Note: the factory settings for set points are 100 mTorr and 200 mTorr for SP1 and SP2, respectively.
To restore these default values, hold

SP1

,

SP2

,

UP

, and

DOWN

simultaneously.

email: sales@frederickscompany.com

tel: +1 215 947 2500

web: frederickscompany.com

mv2a_im rev B

4 of 6

Summary of Contents for MV2A

Page 1: ...2A Thermocouple Controller Instruction Manual The Fredericks Company Televac 2400 Philmont Avenue Huntingdon Valley PA 19006 web www frederickscompany com email sales frederickscompany com tel 1 215 947 2500 fax 1 215 947 7464 ...

Page 2: ... 4 Calibration 5 2 4 1 Setting 1 mTorr Calibration Point 5 2 4 2 Setting 1 Torr Calibration Point 5 2 5 Analog Output 5 3 Part Numbers 6 List of Tables 1 Operating Specifications 2 2 Controller Cable and Sensor Part Numbers 6 List of Figures 1 MV2A Dimensional Drawing 3 2 MV2A Electrical Connection Information 4 3 MV2A Analog Output 5 email sales frederickscompany com tel 1 215 947 2500 web freder...

Page 3: ...s relays and a 0 5 Volt analog output 1 3 Operating Specifications Table 1 Operating Specifications Operating Range 1 10 3 to 20 Torr Communications N A Programmable Set Points 2 Set Point 1 Relay Set Point 2 Relay Analog Output 0 to 5 V Linear Supply Voltage 90 V to 135 V AC 60 Hz Calibration Medium Dry air or nitrogen Overpressure Sensor dependent Analog Output Resolution 12 bits Operating Tempe...

Page 4: ...s all liability for the customer s failure to comply with these requirements READ instructions Read all safety and operating instructions before operating the product RETAIN instructions Retain the Safety and Operating Instructions for future reference HEED warnings Adhere to all warnings on the product and in the operating instructions FOLLOW instructions Follow all operating and maintenance inst...

Page 5: ...e sensor may affect accuracy and decrease the life of the sensor Exposure to oils and other contaminants will decrease the accuracy and decrease the life of the sensor Do not expose the sensor to corrosive gases 2 2 Electrical Information The MV2A has several connectors located on the rear of the unit A connection diagram is provided for convenience Figure 2 MV2A Electrical Connection Information ...

Page 6: ...e holding the set point buttons press the Up and Down arrows until the MV2A reads 1 mTorr 2 4 2 Setting 1 Torr Calibration Point After setting the 1 mTorr calibration use the leak control on the vacuum system to adjust the pressure in the chamber until the reference gauge reads 1 Torr Then press and hold the SP1 and SP2 buttons on the MV2A While holding the set point buttons press the Up and Down ...

Page 7: ... Male 2 2126 042 2A Sensor Stainless Steel 8 VCR Female 2 2126 043 2A Sensor Stainless Steel 4 VCR Male 2 2126 044 2A Sensor Stainless Steel 4 VCR Female 2 2126 045 2A Sensor Stainless Steel 1 33 CF DN16 ISO CF F 2 2126 050 2A Sensor Stainless Steel 2 75 CF DN40 ISO CF F 2 2126 052 3 ft Cable 2 9800 076 10 ft Cable 2 9800 077 20 ft Cable 2 9800 078 35 ft Cable 2 9800 079 50 ft Cable 2 9800 080 Vac...

Reviews: