9.4.5
RS-485 and USB Set Point Commands
Command
Response
Description
RC1
xyy
Read 1E-6 calibration adjustment where x is:
0=negative
1=positive
yy can be 00 to 99
RC2
xyy
Read 1E-5 calibration adjustment
RC3
xyy
Read 1E-4 calibration adjustment
RC4
xyy
Read 1E-3 calibration adjustment
RC5
3
Set analog output format to linear-by-decade
WC1xyy
ppsee
Set 1E-6 calibration adjustment, response is new
pressure reading.
WC2xyy
ppsee
Set 1E-5 calibration adjustment, response is new
pressure reading.
WC3xyy
ppsee
Set 1E-4 calibration adjustment, response is new
pressure reading.
WC4xyy
ppsee
Set 1E-3 calibration adjustment, response is new
pressure reading.
WC5xyy
ppsee
Write analog output adjustment. Return value is
pressure reading.
9.4.6
USB Commands for Setting RS-485 Communications Settings
Command
Response
Description
C1
xxxx
Return serial baud rate, where xxxx is:
0096 = 9600
0192 = 19200
0384 = 38400
0576 = 57600
1152 = 115200
Return new value before changing.
C1xxxx
xxxx
Set serial baud rate, where xxxx is:
0096 = 9600
0192 = 19200
0384 = 38400
0576 = 57600
1152 = 115200
Return new value before changing.
C2
x
Read RS-485 address, where x is 0 to F
C2xx
x
Set RS-485 address, where x is 0 to F
Return new value before changing