MM810A
14
15
Accuracy
(cont'd)
Continuity
Tone Threshold: Between 20 Ω and 300 Ω
Response Time: < 100 μs
AC Current
10 A continuous, 15 A for 30 sec. maximum with 5 minute cool down.
DM-810A • DM-820A • DM-830A • DML-430A
21
Accuracy
(cont’d)
Resistance and Conductance (Conductance on DM-820A, DM-830A, and DML-430A Only)
Range
Accuracy
Typical Open Circuit Voltage
600.0 Ω
± (0.1% + 0.3 Ω)
1.2 VDC
6.000 kΩ
± (0.1% + 0.003 kΩ)
60.00 kΩ
± (0.1% + 0.03 kΩ)
600.0 kΩ
± (0.1% + 0.3 kΩ)
6.000 MΩ
± (0.4% + 0.003 MΩ)
60.00 MΩ
± (1.5% + 0.05 MΩ)
99.99 nS
± (0.8% + 0.1 nS)
1.2 VDC
Continuity
Tone Threshold: Between 20 Ω and 300 Ω
Response Time: < 100 µs
Capacitance
Range
Accuracy
(1)
60.00 nF
± (0.8% + 0.03 nF)
600.0 nF
± (0.8% + 0.3 nF)
6.000 µF
± (1.0% + 0.003 µF)
60.00 µF
± (2.0% + 0.03 µF)
600.0 µF
(2)
± (3.5% + 0.5 µF)
6.000 mF
(2)
± (5.0% + 0.005 mF)
25.00 mF
(2)
± (6.5% + 0.05 mF)
(1)
Accuracies with film capacitor or better
(2)
In manual ranging mode, measurements are not specified below 50.0 µF, 0.54 mF, and 5.4 mF for
600.0 µF, 6.000 mF and 25.00 mF ranges, respectively
Diode Test
Measuring Range: 2.000 V
Test Current (typical): 0.4 mA
Open Circuit Voltage: < 3.5 VDC
Accuracy: 1.0% + 0.001 V
22
Accuracy
(cont’d)
AC Current (AC + DC Current on DM-830A and DML-430A Only)
Range
Accuracy at
50 to 60 Hz
Accuracy at
40 to 1 kHz
Burden Voltage (typical)
(all frequency ranges)
600.0 µA
± (0.6% + 0.3 µA)
± (0.8% + 0.4 µA)
0.08 mV/µA
6000 µA
± (0.6% + 3 µA)
± (0.8% + 4 µA)
60.00 mA
± (0.6% + 0.03 mA)
± (0.8% + 0.04 mA)
2.1 mV/mA
600.0 mA
± (1.0% + 0.3 mA)
± (1.0% + 0.4 mA)
6.000 A
± (0.8% + 0.006 A)
± (0.8% + 0.006 A)
0.02 V/A
10.00 A
± (0.8% + 0.06 A)
± (0.8% + 0.06 A)
10 A continuous, 20 A for 30 sec. maximum with 5 minute cool down.
DC Current
Range
Accuracy at 50 to 60 Hz
Burden Voltage (typical)
(all frequency ranges)
600.0 µA
± (0.2% + 0.4 µA)
0.08 mV/µA
6000 µA
± (0.2% + 4 µA)
60.00 mA
± (0.2% + 0.04 mA)
2.1 mV/mA
600.0 mA
± (0.2% + 0.4 mA)
6.000 A
± (0.2% + 0.004 A
0.02 V/A
10.00 A
± (0.2% + 0.04 A)
10 A continuous, 15 A for 30 sec. maximum (20 A for DML-430A) with 5 minute cool down.
Line Level Frequency
Function
Sensitivity (Sine RMS)
Range
60.00 mV
40 mV
15.00 Hz to 50.00 kHz
600.0 mV
60 mV
15.00 Hz to 50.00 kHz
9.999 V
2.5 V
15.00 Hz to 10.00 kHz
99.99 V
25 V
15.00 Hz to 10.00 kHz
999.9 V
100 V
15.00 Hz to 10.00 kHz
600.0 µA
45 µA
15.00 Hz to 3.000 kHz
6000 µA
600 µA
15.00 Hz to 3.000 kHz
60.00 mA
40 mA
15.00 Hz to 3.000 kHz
600.0 mA
60 mA
15.00 Hz to 3.000 kHz
6.000 A
4 A
15.00 Hz to 3.000 kHz
10.00 A
6 A
15.00 Hz to 3.000 kHz