OPERATION MENU’S
6.1.5
Tube control mode
Tube Control takes place by Touchscreen and TSC-II. Tube Control Mode Menu allows to
define which one of the user interfaces is allowed to control the tube. This is to provide
unauthorized or unwanted recipe selection, or starting and stopping a recipe.
It can be important to disable one (or both) controller interface(s) to prevent any dangerous
situation, like starting a recipe from TSC-II (remote control) at the same time the
maintenance engineer is working on the system.
Figure 6-6 Tube Control Mode Menu
6.1.5.1
Start Recipe Control
Start Recipe Control mode allows 4 levels of authority concerning the tube control by TSC-II
or Touch Screen:
REMOTE (TSC)
Tube control is only possible by using the TSC-II server. In this configuration the Touch
Screen is blocked.
NORMAL (Local + Remote)
Tube control is possible with the Touch Screen as well as the TSC-II server.
DISABLED
Tube control is not possible at all. Not by the Touch Screen and not by the TSC-II.
LOCAL (Touch Screen)
Tube control is only possible by using the Touch screen. In this configuration the TSC-II
server is blocked.
NOTE
Due to safety reasons, aborting a recipe is always possible via the
touch screen or the TSC2.