UM-ALPHA-140821 REV.D
▪
You need to stop recording through TempCentre software if you configured device with both
Loop Recording and
Disable Stop Button.
Alpha TD80 / 80S do not support loop recording.
Logging Interval
The recording and sampling interval for data logger, optional from 1 minute to 2 hours.
Startup Mode
Manual Start
Device needs to be started by pressing
Start/Stop
button for
3
seconds, and 2 additional options
available under this mode, select one between the two.
1.
Start Delay
The time device needs to wait before starting recording after manually activated, optional from 1
minute to 30 days.
2.
Condition Start (Temperature Start)
This function allows loggers to start recording when environment temperature meets pre-set
temperature condition, after setting a logger to condition start, you need to start logger to measure
environment temperature by pressing
Start/Stop
button for 3 seconds, then when environment
temperature meets the start condition, logger would start recording automatically.
There are three temperature start conditions available:
Over,
device would start recording when temperature goes over the temperature value preset.
Under,
device would start recording when temperature goes under the temperature value
preset.
Between,
device would start recording when temperature falls in between the temperature
range preset.
Auto Start
Device automatically starts recording on the specific date/time preset, no need to press
Start/Stop
button after configuration.