You  need  to  stop  recording  through  TempCentre  software  if  you  configured  device  with  both 

Loop Recording and 

Disable Stop Button.

 

 

Alpha TD80 / 80S do not support loop recording. 

 

 

Logging Interval 

The recording and sampling interval for data logger, optional from 1 minute to 2 hours. 

 

Startup Mode 

 

Manual Start 

Device needs to be started by pressing 

Start/Stop

 button for 

3

 seconds, and 2 additional options 

available under this mode, select one between the two. 

 

1.

 

Start Delay 

The time device needs to wait before starting recording after manually activated, optional from 1 

minute to 30 days. 

 

2.

 

Condition Start (Temperature Start) 

 

This function allows loggers to start recording when environment temperature meets pre-set 

temperature condition, after setting a logger to condition start, you need to start logger to measure 

environment temperature by pressing 

Start/Stop

 button for 3 seconds, then when environment 

temperature meets the start condition, logger would start recording automatically. 

 

There are three temperature start conditions available: 

 

 

Over, 

device would start recording when temperature goes over the temperature value preset.

 

 

Under, 

device would start recording when temperature goes under the temperature value 

preset.

 

 

Between, 

device would start recording when temperature falls in between the temperature 

range preset.

 

 

Auto Start 

Device automatically starts recording on the specific date/time preset, no need to press

 Start/Stop

 

button after configuration. 

 

 

 

Summary of Contents for Alpha T30

Page 1: ...2014 TempSen Electronics Company USER MANUAL EN TempSen Alpha T30 TH30 T50X TD80 TD80S TUX UM ALPHA 140821 REV D Updated 07 06 2015...

Page 2: ...ngs 12 PDF Report Settings 12 Recording 13 Start recording 13 Manual Start 13 Auto Start 13 In Recording 14 Overall Alarm Status Default Screen 14 Overall Limit Violation Events Default Screen 14 Moni...

Page 3: ...D Sensors 23 Response Time Temperature 23 Response Time Test 23 MKT 23 Why is MKT important in the life of pharmaceuticals and perishable goods 23 Time Error 24 Additional Information 24 Warranty 24...

Page 4: ...UM ALPHA 140821 REV D Products Overview 1 USB plug 2 LCD Screen 3 View Mark button 4 Start Stop button 5 External Sensor Probe1 1 Only available on products supporting external sensors...

Page 5: ...readings Measurement Range 30 C 70 C 22 F 158 F 30 C 70 C 22 F 158 F 0 RH 100 RH 50 C 70 C1 58 F 158 F 80 C 70 C 112 F 158 F 200 C 250 C1 328 F 482 F Type1 70 C 250 C Type2 100 C 150 C Type3 200 C 15...

Page 6: ...with embedded raw data Dimensions 107mm L x 48mm W x 12mm H 107mm L x 56mm W x 18mm H 107mm L x 48mm W x 12mm H Weight Approx 65g Approx 65g Approx 75g Approx 62 80g Approx 65g exclude external probe...

Page 7: ...already signed up download now button in dialog box then software would begin to download in couple of minutes Installing and using TempCentre software Please refer to TempCentre User Manual for insta...

Page 8: ...FAST mode to skip PDF report generation and start configuration right away Device will quit FAST mode after 10 seconds if there is no USB connection following Applying Configuration LCD displays ConF...

Page 9: ...d Start Stop Logger stops recording until memory is full Loop Endless Recording continues even memory is full under this mode new data point overrides old data point in memory logger provides monitori...

Page 10: ...t This function allows loggers to start recording when environment temperature meets pre set temperature condition after setting a logger to condition start you need to start logger to measure environ...

Page 11: ...nabled you could restart Alpha data logger for next round of monitoring with the same recording parameters of previous recording by pressing Start Stop button no need to program the logger again rPt f...

Page 12: ...ltiple alarm ranges Alarm Type Single Event Alarm could only be triggered when data over under alarm limit for a consecutive time not less than alarm delay Cumulative Alarm Alarm could be triggered on...

Page 13: ...cy of countdown refreshes as per delay time remains for example when there is 2 hours left delay time refreshes each hour 30 minutes left LCD refreshes each 1 minute In this case start delay time is 3...

Page 14: ...arm Triggered A Cross appears in upper left corner of screen when any pre set alarm is triggered Overall Limit Violation Events Default Screen Limit Violation Indication High Low Alarm Setting High an...

Page 15: ...ximam Value Minimum Value Average Value Information of Each Alarm Zone Examples Alarm Zone A1 Duration time 0 minute Limit violation event of this zone No Status of this zone OK Alarm Zone A2 Duration...

Page 16: ...k is about to be placed To successfully add a time mark please hold the View Mark button throughout the whole process a maximum of 9 marks are available for adding A timestamp would be written in logg...

Page 17: ...ger to PC open TempCentre software and click Download in toolbar then TempCentre would search connect logger and download data automatically Please remember to save data to data base of TempCentre by...

Page 18: ...www coldchaincloud com to enable shipment build and worldwide data sharing alarm notification between shipper receiver and QA management personnel Authenticity and integrity of PDF reports created by...

Page 19: ...UM ALPHA 140821 REV D PDF Report Sample Please go to www coldchaincloud com to get latest information about Cold Chain CloudTM online data storage and sharing service...

Page 20: ...is shown in days on LCD battery life will minus one day very calendar day after first configuration Battery life for single use model Alpha logger will start to deduct after recording started Remaini...

Page 21: ...ternal probe to install and remove probe LCD displays nC No Connection when external sensor probe is removed please connect probe back to device to proceed IMPORTANT Removal of external probe from dev...

Page 22: ...using screws double sided tapes and magnet Please install the logger in bracket from bottom up Bracket is not provided with Alpha TD80 TD80S Calibration Management Enable calibration expiry notificati...

Page 23: ...perature is a simplified way of expressing the overall effect of temperature fluctuations during storage or transit of perishable goods Calculation E R ln 1 2 E activation energy standard value 83 144...

Page 24: ...t no charge using new parts or parts equivalent to new in performance and reliability 2 exchange the new product with a product is new or equivalent to new in performance and reliability and is at lea...

Page 25: ...on ports c to damage caused with non TempSen products d to damage caused by accident abuse misuse flood fire earthquake or other external causes e to damage caused by operating the product outside the...

Reviews: