UM-ALPHA-140821 REV.D
Alarm Settings
Alarm Mode
No alarm setting
High & Low alarm, 2 alarm ranges determined by 2 alarm limits.
Multiple Alarm ranges, 5 alarm ranges determined by 5 alarm limits.
High range: A1, A2, A3
Ideal range: A4 (No alarm will be triggered when data falls within this range )
Low range: A5, A6
▪
Alpha TH30 does not support multiple alarm ranges.
Alarm Type
Single Event
Alarm could only be triggered when data over/under alarm limit for a consecutive time not less than
alarm delay.
Cumulative Alarm
Alarm could be triggered once the cumulatively total time data within alarm range is not less than
alarm delay.
PDF Report Settings
PDF Report Title
This report title would be used for PDF report created automatically by data logger.
Time Base
Select Time Zone and enable/disable Day Light Saving time (DST) for PDF report.
Language
Select language for PDF report.