UM-ALPHA-140821 REV.D
LCD displays
“
Err
” for configuration failure, please re-configure device with correct
parameters.
Configuration Options
Trip Number
A unique 10 digits alphanumeric number automatically created for identifying each monitoring task.
Trip Description
A brief note entered by user for describing monitoring task and showing on data report.
Logging Mode
Standard (Start/Stop)
Logger stops recording until memory is full.
Loop (Endless)
Recording continues even memory is full under this mode, new data point overrides old data point in
memory, logger provides monitoring and alarm statistics over entire monitoring course. Data points
available will be the latest full memory readings.
“
”
displays when Loop recording is enabled.
IMPORTANT
Configuration to data logger would erase all data recorded, so please make sure you
have save the data before configuration.