Safe‐T‐Vue® 10 Instructions

 

Preparation 

Refrigerate Safe‐T‐Vue® for a minimum of 24 hours before activating. 

•Use a refrigerated cold pack beneath blood products to keep them 

cold during the dispensing process (recommended). 

 

Handling 

Hold the Safe‐T‐Vue® indicator by the outer 

edges. 

•Do not hold or touch the temperature sensitive  

indicator areas, as your hands can pre‐warm 

the indicator and affect its accuracy. 

 

Application 

1. 

Apply the Safe‐T‐Vue® indicator to the blood bag 

•When the blood is requisitioned and dispensed 

from the blood bank, OR 

•When the blood is received into your blood 

bank inventory

Color Change Temperature Indication 

The white dot will begin to gradually fill in  

with red dots at approximately 9°C. 

•The appearance of small red dots (typically 

starting around 9°C) is an indication that the  

blood should be cooled or re‐refrigerated 

immediately. 

•The temperature‐sensitive dot on Safe‐T‐Vue® 

10 will turn from white to completely red 

when the blood product temperature reaches 

10°C. 

•Extensive testing has shown that the color  

change from white to completely red should 

be clear and unambiguous if Safe‐T‐Vue has 

been handled properly. 

 

Validation 

 
 
 
 

 

< 9° C 

 
 
 
 
 

≈ 9.5° C 

 
 
 
 

 

≥ 10° C

2. 

Remove the adhesive backing and apply  

Safe‐T‐Vue® to the lower area of the blood bag,  

where there is the largest blood volume. 

3. 

Be careful not to touch the center indicator areas.

 

 

Activation 

1. 

Pull back the foil to expose the indicator’s active  

areas. Again, be careful not to touch the center  

indicator areas. 

2. 

Activate the indicator by folding the rounds  

together and snap shut.

 

3. 

The blood is ready for transport to destination,  

or can be placed back into the refrigerator until  

needed. 

 

A detailed Validation Protocol, is available for download in MSWord format at 

www.williamlabs.com/support. Many factors such as ambient temperature, 

bag size, indicator handling, and air flow can influence the results of your 

validation testing. To account for these variables, the validation protocol can 

be edited and modified to meet the requirements of your facility. Once 

validated, you may continue to use the indicators according to your validated 

procedure. Contact customer service at 973‐630‐6050 for support.

 

 

Recommended equipment and processes: 

An electronic thermometer and thermister probes (±0.1°C stated accuracy) 

placed inside simulated blood bags are recommended for validation of Safe‐T‐

Vue® 10. Avoid surface measurement such as infrared thermometers (±1.5°C 
typical stated accuracy) or glass thermometers with contact only on the 
surface of the bag. 

•Simulated blood products may be prepared using the appropriate volume of 

water with glycerol‐ see 

www.williamlabs.com 

for simulated blood product 

recipes. 

Symbols Glossary: Refer to ISO 15223-1 Standard 

Symbol 

Meaning 

Symbol 

Meaning 

 

Manufacturer 

 

Caution: Consult the Instructions 

 

Do not re‐use 

 

Use By Date 

 

Catalogue or Part Number 

  

Batch Code 

 