A detailed 

Validation Protocol

, is available for download in MSWord 

format at www.williamlabs.com/support. Many factors such as ambient 
temperature, bag size, indicator handling, and air flow can influence the 
results of your validation testing. To account for these variables, the 
validation protocol can be edited and modified to meet the requirements 
of your facility. Once validated, you may continue to use the indicators 
according to your validated procedure. Contact customer service at 973-
630-6050 for support. 

 

Recommended equipment and processes:

 

 

An electronic thermometer and thermister probes (±0.1°C stated 

accuracy) placed inside simulated blood bags are recommended for 
validation of Safe-T-Vue® 6. Avoid surface measurement such as infrared 
thermometers (±1.5°C typical stated accuracy) or glass thermometers with 
contact only on the surface of the bag. 

• 

Simulated blood products may be prepared using the appropriate volume 

of water with glycerol- see 

www.williamlabs.com 

for simulated blood 

product recipes. 

Symbols Glossary: Refer to ISO 15223-1 Standard 

Symbol 

Meaning 

Symbol 

Meaning 

 

Manufacturer 

 

Caution: Consult the Instructions 

 

Do not re-use 

 

Use By Date 

 

Catalogue or Part Number 

  

Batch Code 

 