A detailed
Validation Protocol
, is available for download in MSWord
format at www.williamlabs.com/support. Many factors such as ambient
temperature, bag size, indicator handling, and air flow can influence the
results of your validation testing. To account for these variables, the
validation protocol can be edited and modified to meet the requirements
of your facility. Once validated, you may continue to use the indicators
according to your validated procedure. Contact customer service at 973-
630-6050 for support.
Recommended equipment and processes:
•
An electronic thermometer and thermister probes (±0.1°C stated
accuracy) placed inside simulated blood bags are recommended for
validation of Safe-T-Vue® 6. Avoid surface measurement such as infrared
thermometers (±1.5°C typical stated accuracy) or glass thermometers with
contact only on the surface of the bag.
•
Simulated blood products may be prepared using the appropriate volume
of water with glycerol- see
www.williamlabs.com
for simulated blood
product recipes.
Symbols Glossary: Refer to ISO 15223-1 Standard
Symbol
Meaning
Symbol
Meaning
Manufacturer
Caution: Consult the Instructions
Do not re-use
Use By Date
Catalogue or Part Number
Batch Code